ea0013p117 | Comparative | SFEBES2007

Performance characteristics of an intact PTH recently developed for the Abbott Architect immunoassay analyser

Smith Karen , Wallace Mike , Louden Ian

Concerns on the noncomparibility of intact parathyroid hormone (iPTH) methods have recently been voiced (Cantor T et al; Clin Chem 2006, 52, 1771–6). It is therefore important that any new iPTH immunoassays are thoroughly validated before routine use. We describe performance characteristics for an iPTH immunoassay on the ARCHITECT 8200i analyser as compared with iPTH measurement on the Diasorin LIAISON platform. Potassium EDTA plasma samples (n=80) we...